Chlorine dioxide Sensor AS2 / AS3
Sensor for the measurement of chlorine dioxide with open measuring cell.
Product description:
Measurement: Chlorine dioxide
Calibration: at the controller, via analytical determination by DPD-1 method
Interferences: ozone, chlorine, chlorite are registered with less than 2%
pH range: 1 ... 9,
Pressure range: 0 ... 8 bar,
Temperature range: 0 ... 50 °C (AS2), 0 ... 70 °C (AS3),
Automatic temperature compensation
Response time: T90 approx. 30 s,
Flow rate: approx. 30 l/h,
Shaft length: 175 mm, 195 mm, mA-version 220 mm,
Material: PVC-U (AS2), PEEK (AS3)
Areas of application:
• Fresh water, especially drinking water, up to max. 70 °C
Scope of supply:
• AS2, AS3: sensor, electrolyte hull, electrolyte